Technical data
40
9 Technical data (to DIN 58966)
Low-temperature thermostats
RE 204 RE 206 RE 207 RE 210 RE 212 RE 220
Operating temperature range °C -10 to 200 -20 to 200 -35 to 200 -40 to 200 -30 to 200 -30 to 200
Ambient temperature range °C 5...40
(!Section 5.1)
Setting resolution °C 0.01
Indication resolution °C 0.05
Indication accuracy ±0.2 °C additive re-calibration
Temperature control ± °C 0.01 0.01 0.02 0.02 0.02 0.02
Safety features FL
Additional functions 2-line LCD-display
autom. compressor control autom. compressor control + cooling adjustments
Heater
power
230 V; 50/ 60 Hz
115 V; 60 Hz
100 V; 50/60 Hz
kW 2.25
1.3
1.0
Cooling capacity (eff.)
@ with ethanol at 20 °C
ambient temperature
20°C
0°C
-20°C
-30°C
kW 0.18
0.12
0.05*
----
0.20
0.15
0.05
----
0.30
0.22
0.10
0.06
0.50
0.42
0.27
0.14
0.04
0.30
0.23
0.13
0.04
0.35
0.25
0.10
0.04
Pump type pressure pump with choice of 5 output steps
Max. discharge pressure bar 0.4
Max. flow rate l/min 17
Pump connections mm nipples 13 mm dia.
Max. bath volume l 3 to 4.5 4 to 6 4 to 6
6,5...9,5
9 to 12 14 to 20
Bath opening (W x D) mm 130x105 150x130 150x130
200x200
200x200 300x350
Bath depth mm 160 160 160
160
200 160
Usable depth mm 140 140 140
140
180 140
Height top edge of bath mm 363 396 396
441
441 441
Overall size (W x D x H) mm 180x320x524 200x400x557 200x400x557
280x400x
602
250x400x602 350x530x602
Weight kg 21 24 26
30
30 34
Power
con-
sumption
230V; 50/ 60Hz
115V; 60 Hz
100V; 50/60 Hz
kW 2.3
1.4
1.1
2.3
1.4
1.1
2.3
1.4
1.1
2.3
1.4
1.1
2.3
1.4
1.1
2.3
1.4
1.2
Ref.-no. 230 V; 50/60 Hz
230 V; 50 Hz
230 V; 60 Hz
115 V; 60 Hz
100 V; 50/60 Hz
208-230V; 60Hz
LCK 0862
-----
-----
LCK 4862
LCK 6862
-----
LCK 0865
-----
-----
LCK 4865
LCK 6865
-----
-----
LCK 1868
LCK 2868
LCK 4868
LCK 6868
-----
-----
LCK 1883
-----
LCK 4883
-----
LCK 8883
-----
LCK 1871
LCK 2871
LCK 4871
LCK 6871
-----
----
LCK 1874
LCK 2874
LCK 4874
LCK 6874
-----
*@ -10°C
FL: suitable for flammable and non-flammable liquids
Section 7.2 last item
at pump output step 5
Units to EU Directive 89/336/EWG ( EMC ) and 73/ 23/ EWG (low-voltage) with CE-mark
We reserve the right to make technical alterations!