6.4.4.2 Setting the reset time of the PID-controller
and 3x
Directly from level 0 (setpoint selection), press the key
combination on the left, until the PARA function is reached,
then
and 1x
switch to level 2 and move forward with the keys on the left.
The display shows the current bath temperature, Tn and the
current setting. To alter the setting, press the key on the left.
! Possible adjustment from 5 to 60 s
The display is flashing (approx. 4 s).
or
During this time, start to set the required value with one of
the two keys.
Forward with the key to "Selection of the interface" or
with key back to "Setting the proportional band".
The control parameters are pre-set to suit the unit type.
Normally no change is required. Some adjustment is necessary only when using
Silicone oil and with very stringent demands on short-term stability. If there are control
fluctuations, increase the values for Xp and Tn. If the setpoint is not reached
!
select
smaller values. The derivative time Tv (D-part) is altered automatically through a fixed
factor to Tn.
(Standard settings of control parameters and pump Section 9. Technical data)