Starting up
15.05.01
YACE0060
33
6.5.2 Low-level protection and testing
$
$$
$

Double signal tone sounds, if the bath liquid falls so
much that the heater is no longer covered with liquid
completely
1. The display shows LEVEL (low-level) and the fault
triangle is flashing
!
heating is switched off on both poles,
!
pump and refrigeration system are switched off by
the electronics.
2. Top up the bath Section 5.2 and reset with the key.
If necessary repeat several times in case that the pump does not purge
immediately.
Testing at regular intervals by lowering the bath level. Place a hose on the pump
connector and pump some of the bath liquid into a suitable container.
Step 1 - 2 must follow.
Bath temperature during this test not below 0°C or higher than 50°C, otherwise
danger of burn injuries !
If there is any irregularity when testing the safety devices, switch off the unit
immediately and pull out the mains plug !
Have the unit checked by the LAUDA service or the local service organisation!