Starting up
32
The position of the potentiometer is decisive for the setting. The display is just a help
for the setting.
Setting is possible only up to a upper limit of the operating temperature range + 5 °C.
$
$$
$

When the bath temperature arises above the
overtemperature switch-off point.
1. Double signal tone sounds
2. The display shows the indication for overtemperature
(TEMP) the fault triangle is flashing
! heating is switched off on both poles,
! pump and refrigeration system are switched off by the
electronics.
Rectify the cause of the fault.
Wait until the bath temperature has cooled down below the
switch-off point or set the switch-off point at a higher value.
When the display shows TEMP.
Reset with the key
Before the unit is running unattended for longer periods overtemperature protection
should be tested. Therefore
Turn the potentiometer slowly anticlockwise.
!
The unit
must switch off at the bath temperature.
Step 1 - 2 (see above) must follow.
Set the overtemperature switch-off point again above the
bath temperature and wait until the indication TEMP
appears on the display, then
reset with the key.