EasyManua.ls Logo

Lauda Ecoline RE 204 - Write Commands (Data Commands to the Thermostat)

Lauda Ecoline RE 204
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Unit description
10
3.6.3 Write commands (data commands to the thermostat)
Command Explanation
OUT_SP_00_XXX.XX Setpoint transfer with up to 3 places before the decimal point and up
to 2 places behind
OUT_SP_01_XXX Pump output step 1, 2, 3, 4, or 5; 0 = Stop (Standby)
OUT_SP_02_XXX Refrigerator 0 = OFF / 1 = ON 100% / 2 = 50% (reduced capacity
RE 207, 210, 212, 220) / 3 = automatic operation
OUT_PAR_00_XXX.XX Setting of control parameter Xp for controller (0.5 to 9,9°C)
OUT_PAR_01_XXX Setting of control parameter Tn (5 to 60 sec)
OUT_MODE_00_X Keys: 0 = free / 1 = inhibited
START Switches thermostat on (from standby).
STOP Switches thermostat to standby (pump, heating, refrigerator off).
For "_" use also " " (blank character).
Response from thermostat "OK" or in case of error "ERR_X" (RS485 interface e.g.
"A015_OK" or in case of error "A015_ERR_X").
Permitted data formats:
-XXX.XX -XXX.X -XXX. -XXX XXX.XX XXX.X XXX. XXX
-XX.XX -XX.X -XX. -XX XX.XX XX.X XX. XX
-X.XX -X.X -X. -X X.XX X.X X. X
-.XX -.X .XX .X

Related product manuals