Starting up
27.09.02
YACE0060
31
6. Switch back to level 0 or
7. with key back to "Setting the transfer rate
Example
a) Insert a suitable thermometer into the bath (long enough and far enough).
b) Remove the external consumers and turn the setting knob of the pump outflows to the
right side.
c) Set the setpoint to a temperature where you use to work (e.g. set the setpoint to 45°C
( Section 6.3.)
d) Wait until the actual bath temperature has reached the setpoint temperature of 45°C
and until the indication on the reference thermometer does not change any more.
e) The reference thermometer shows e.g. 44,8 °C.
f) Select CAL on the display and go forward as mentioned under point 1-7. The actual
bath temperature switches from 45°C to 44,8°C and the unit starts to heat up until the
actual bath temperature has reached 45°C. (
!
the reference thermometer should also
indicate 45°C).
6.5 Warning and safety functions
6.5.1 Overtemperature protection and testing
The units are designed for operation with non-flammable and flammable liquids to
EN 61010-2-010
Set the overtemperature switch-off point. Recommended
setting 5°C above required bath temperature.
Not higher than 25 °C below the firepoint of
the bath liquid ( Section 5.3.).
The actual switch-off point is indicated on the display, e.g.
max. 110°C
When the switch-off point is being adjusted by more than 2°C ! display shows MAX