The test run takes between 30 minutes and a maximum of 3 hours,
depending on the application. During the run, the temperature of the heat
transfer liquid deviates a maximum of ±15 Kelvin from the defined setpoint.
After completion of the test run, the newly determined values of the control
parameters are accepted and stored in the [Control parameter] menu. The
old control parameters are overwritten.
1. Press the [Enter key] to open the menu.
2.
Select the menu items Setup Control Self adaptation.
A submenu opens.
3. Start self adaptation here:
n [Start self adaptation]
The self adaptation run begins.
You can stop the self adaptation process prematurely by
pressing [Cancel self adaptation].
n Set point
Specify the temperature setpoint T
set
here.
Take the maximum temperature fluctuations of up to max-
imum ±15 Kelvin into consideration in the process.
Set [Tih] and [Til] accordingly.
n Determination
With the Only internal
command, you determine the internal
control parameters.
With the Internal and external
command, you determine both
internal and external control parameters.
n Control parameter
Here you can view the control parameters.
n Status
Shows the current phase of the program for determining the
parameters.
WAIT - Self adaptation running
END - Self adaptation ended
SEND - Parameters are transmitted
STOPPED - Self adaptation stopped
6.6 Set the setpoint oset
It is possible to apply an oset value to the temperature measured by an
external temperature probe and then process this temperature as a target
value. The setpoint of the heat transfer liquid in the device can therefore be
set, for example, to 15 K below the temperature in the consumer, measured
by the external temperature probe.
1. Press the [Enter key] to open the menu.
2.
Select the menu items Setup Control Setpoint oset.
Start self adaptation
Navigating to the settings
V6Integral Process Thermostats and High-Temperature Thermostats94 / 198

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lauda IN 130 T and is the answer not in the manual?

Lauda IN 130 T Specifications

General IconGeneral
BrandLauda
ModelIN 130 T
CategoryThermostat
LanguageEnglish

Related product manuals