Please note:
Water may not be used with any Integral XTs over the entire working
temperature range.
Table 6: Heat transfer liquid order numbers
Designation
Container size
Catalog number
5 L 10 L 20 L
Kryo 95 LZB 130 LZB 230 LZB 330
Kryo 71 LZB 131 LZB 231 LZB 331
Kryo 70 LZB 127 LZB 227 LZB 327
Kryo 65 LZB 118 LZB 218 LZB 318
Kryo 51 LZB 121 LZB 221 LZB 321
Kryo 30 LZB 109 LZB 209 LZB 309
Kryo 20 LZB 116 LZB 216 LZB 316
Aqua 90 LZB 120 LZB 220 LZB 320
Ultra 350 LZB 107 --- ---
Ultra 301 LZB 153 LZB 253 LZB 353
4.5 Cooling water requirements
This section is relevant for:
n Water-cooled devices
NOTICE!
Risk of cooling circuit leaking due to corrosion
Device damage
Do not use corrosive cooling water.
Cooling water is subject to specific purity requirements. A suitable procedure
must be employed to purify the cooling water in line with the contamination
in the water and maintain the water quality. Unsuitable cooling water may
cause the condenser and the entire cooling water circuit to become blocked
or damaged, or start to leak. The entire cooling circuit and cooling water
circuit may sustain extensive consequential damage as a result.
Requirements
V6Integral Process Thermostats and High-Temperature Thermostats42 / 198