4 Before starting up
4.1 Install device
WARNING!
Rolling away or overturning of the device due to incorrect han-
dling
Impact, crushing
Do not tilt the device.
Position the device on an even, non-slip surface with a
sucient load carrying capacity.
Actuate the castor brake when setting up the device.
Do not place heavy parts on the device.
WARNING!
Degradation of overflow or drain
Electric shock
Overflow hose and drainage hose must each slope down-
ward separately all the way into a collecting vessel.
WARNING!
Danger of overpressure from high ambient temperature
Injury, escape of refrigerant
Note the permitted ambient temperature.
CAUTION!
Risk of heat transfer liquid escaping
Scalding, cold burns
The temperature and media resistance of the hoses used for
the device overflow must be suitable for the application.
Installation conditions:
n Irritant vapors may develop, depending on the heat transfer liquid and
operating temperature used. Always ensure that the vapors are ade-
quately extracted.
n Note the electromagnetic compatibility (EMC) requirements of the
device Ä Chapter 1.3 “EMC requirements” on page 10.
n Do not cover the ventilation openings of the device.
n Further conditions for installation apply to the devices. These are speci-
fied in the technical data Ä Chapter 15.1 “General data” on page 161.
V6Integral Process Thermostats and High-Temperature Thermostats34 / 198

Table of Contents