6.4 Setting Tmax
WARNING!
Overheating of heat transfer liquid due to incorrect entry of
overtemperature switch-o point T
max
for the hydraulic circuit
Fire
Set the overtemperature switch-o point for the hydraulic
circuit to 5 K above the upper limit of the temperature
range for your application. But not above the upper limit of
the working temperature range of the heat transfer liquid.
The following information is only relevant for the Integral T device:
WARNING!
Overheating of heat transfer liquid due to incorrect entry of
overtemperature switch-o point T
max
for the hydraulic circuit
Fire
In the menu of the device, set the heat transfer liquid used
in the device. Set the overtemperature switch-o point
below the flash point of the heat transfer liquid.
The following information is only relevant for the Integral XT device:
WARNING!
Overheating of heat transfer liquid due to incorrect entry
of overtemperature switch-o point T
maxTank
for the expansion
vessel
Fire
In the menu of the device, set the heat transfer liquid used
in the device. Set the overtemperature switch-o point
T
maxTank
below the flash point of the heat transfer liquid.