ID Function Unit Command
80 Continue programmer (after pause) [–] RMP_CONT
81 End programmer [–] RMP_STOP
4.7.8 Error messages
The error messages of the Ethernet interface are described in the following.
The string ERR_X or ERR_XX is output after an incorrect command.
Error Description
ERR_2 Incorrect entry (for example, buer overflow)
ERR_3 Wrong command
ERR_5 Syntax error in value
ERR_6 Impermissible value
ERR_8 Module or value not available
ERR_30 Programmer, all segments occupied
ERR_31
Not possible to specify a set point value. The "Setpoint oset"
function is active.
ERR_32
TiH £ TiL
ERR_33 External sensor missing
ERR_34 Analog value not present
ERR_35 Automatically configured
ERR_36
Not possible to specify set point, programmer is running or
has been paused
ERR_37
Not possible to start programmer (analog set point value input
is ON)
ERR_38
The operator does not have the privileges to execute the
command. Another control section has exclusive operating
rights, which means that writing via this interface is not per-
mitted.
ERR_39 Operation is not permitted. Safe Mode is active.
ERR_40 Operation is not permitted. Safe Mode is switched o.
ERR_41
Operation is not permitted. Constant temperature equipment
is in error status.
V6 Integral Process Thermostats and High-Temperature Thermostats 61 / 198

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lauda IN 130 T and is the answer not in the manual?

Lauda IN 130 T Specifications

General IconGeneral
BrandLauda
ModelIN 130 T
CategoryThermostat
LanguageEnglish

Related product manuals