Integral T process thermostats
24 Preparations YAWE 0026 / 21/01/2019
With external circuits at a higher level, even with a closed circuit, the external
volume may drain down when the pump is stopped and air enters the thermostating
circuit; this may result in overflowing of the storage bath!
Use only hydraulic closed consumers.
Always ensure maximum flow cross-section in the external circuit (nipples, tubing, external
apparatus). This results in increased flow rate and therefore better thermostating.
Protect tubing against slipping off by fitting hose clips!
The equipment is designed for operation with non-flammable and flammable liquids
in accordance with DIN EN 61010-2-010.
Flammable heat transfer liquids may only be used below the flash point.
6.3 Emptying
Place a container underneath the drain cock on
the back of the equipment. Switch off the
equipment!
Observe the appropriate regulation when
disposing of used heat transfer liquid.
Close the drain cock!
Do not drain the heat transfer liquid when it is hot or at temperatures below 0 °C!
Drain cock

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lauda Integral T 7000 W and is the answer not in the manual?

Lauda Integral T 7000 W Specifications

General IconGeneral
BrandLauda
ModelIntegral T 7000 W
CategoryThermostat
LanguageEnglish

Related product manuals