Integral T process thermostats
21/01/2019 / YAWE 0026 Starting up 55
7.6.4.3 Outflow temperature limit
High limit (TiH)
The temperature set here limits the outflow
temperature. Especially with external control this
limitation avoids the undesirable permanent switch-off
through the safety circuit (overtemperature) during the
start-up phase.
Low limit (TiL)
The temperature set here limits the outflow
temperature against excessively low values, e.g. in
order to prevent freezing up.
Input of a value for TiH which is below the value for TiL
causes to appear “END Err” as error message at menu
end.
The values for TiH and TiL are entered only after
leaving the submenu.