Integral T process thermostats
74 Starting up YAWE 0026 / 21/01/2019
With external control switched on:
Pressing scrolls from REG . . to the input for the
control parameters.
With internal control (CON I)
With internal control (CON I) select here the
proportional band Xp for the outflow controller. Input
range 0.1 10 °C. Useful values are 2 7 °C,
depending on external circuit and heat transfer liquid. If
the value is too small (e.g. 2 °C) this may result in
control oscillations.
If the value is too large (e.g. 8 °C) the compensation of
disturbances may be worse and more sluggish.