WaveJet 300A Series
16 917976 Rev A
4. LIMITATIONS ON REMOTE FUNCTION
The following limitations are imposed when Remote Control is selected.
• Delimiters permit the following settings.
Table 4-1. Delimiters
State Delimiter (GPIB and TCP/IP) Delimiter (USB)
Recv EOI CR
Send LF+EOI LF+CR
• Panel Lock mode is entered.
• The setting is changed to the following once Remote Control mode is entered.
Table 4-2. Setup in Remote Control mode
Item Setup Panel
Help display Off -
Offset Setting Division Utilities/Config. 1/2
Power Off Never Utilities/Config. 1/2/Power
Management
Panel Lock Lock Utilities/Config. 2/2
Volts/div Coarse Channel Menu 2/2
Zoom Off ZOOM
Line Number (TV Standard = NTSC) 263/262 Trigger Menu/Type = TV/TV Setting
Line Number (TV Standard = PAL) 625 Trigger Menu/Type = TV/TV Setting