EasyManua.ls Logo

LeCroy WaveJet 300A series - Trigger System Features

LeCroy WaveJet 300A series
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WaveJet 300A Remote Control Manual
917976 Rev A 19
is identified in <mode>.
REMARKS WJ 324, WJ314, WJ322 and WJ312 set the bandwidth limi
t
er
to 20 MHz when these scopes receive "BWL 100M".
Status *CLS
Command
DESCRIPTION The *CLS command clears the entire status data
register.
COMMAND SYNTAX *CLS
QUERY SYNTAX
RESPONSE FORMAT
Cursor CMSR?
Query
DESCRIPTION The CMSR? query returns the currently displayed curso
r
measurements.
COMMAND SYNTAX
QUERY SYNTAX CMSR?
RESPONSE FORMAT <measure1>, <measure2>, <measure3>, <measure4>,
<measure5>, <measure6>, <measure7>
• Current cursor measurements are displayed in
<measure1> to <measure7> in
<NR3 Numeric Response Data> format.
• When the cursor mode is disabled or the data cannot
be measured, the query returns +9.91E+37.
Seven types of cursor measurement results are output
,
respectively separated by a comma, regardless of
whether cursor mode has been selected. For items not
covered by the measurement, the query returns
+9.91E+37.
• The table below shows the relationship between
<measure1> to <measure7> and
cursor mode.
<measure1>:= CH1
<measure2>:= CH2
<measure3>:= CH3
<measure4>:= CH4
<measure5>:= MATH
<measure6>:= delta t
<measure7>:= 1 / delta t
When MATH is FFT, M1:TRA is ON and CURM is VATH, CMSR?
returns frequency at the cursor of the FFT waveform for
<measure7>.
REMARKS When calculating the cursor measurement values of the
YT display Amplitude, Value at
C
ursor, and XY display
ch Amplitude in the model WJ352/332/322/312,
<measure3> and <measure4> return "+9.91E+37".