Parameter Description Definition Notes
Delta delay
delay: Computes time between
50% level of two sour
Time between midpoint Standard parameter.
ces. transition of two
sources
Dperiod@level
Adjacent cycle deviation (cycle-to-
cycle jitter) of each cycle in a
waveform
Reference levels and edge-
transition polarity can be
selected. Hysteresis argument
used to discriminate levels from
noise in data.
Available with JTA2 and XMAP
options.
Dtime@level
t at level: Computes transition
between selected levels or sources.
Time between
transition levels of two
sources, or from trigger
to transition level of a
ource
Reference levels and edge-
transition polarity can be
selected. Hysteresis argument
used to discriminate levels from
noise in data. single s
This measurement
gives the time of the
source 2 edge minus
the time of the source 1
edge.
Standard parameter.
Duration
For single sweep waveforms, dur is
: time from previous
t's to current segment's
ted waveform's
Time from first to last Standard parameter.
0; for sequence waveforms: time
from first to last segment's trigger;
for single segments of sequence
acquisition: for
average, histogram or
sequence waveforms
waveforms
segmen
trigger; for waveforms produced by
a history function: time from first to
last accumula
trigger.
Duty@level
e
scriminate levels from
Available with JTA2 and XMAP
Percent of period for which data ar
above or below a specified level.
Reference levels and edge-
transition polarity can be
selected. Hysteresis argument
used to di
noise in data.
options.
Duty cycle
rcentage of
period.
riod Standard Duty cycle: Width as pe width/pe parameter.
Dwidth@level
Difference of adjacent width above
or below a specified level.
t
d XMAP
Reference levels and edg
transition polarity can be
e-
selected. Hysteresis argumen
used to discriminate levels from
noise in data.
Available with JTA2 an
options.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LeCroy WaveRunner XI Series and is the answer not in the manual?

LeCroy WaveRunner XI Series Specifications

General IconGeneral
BrandLeCroy
ModelWaveRunner XI Series
CategoryTest Equipment
LanguageEnglish

Related product manuals