Use the procedure below to shut down the pump:
1. If shutting the pump down prior to a period of storage, remove any1.
process gases by running on a gas ballast for at least one hour.
2. Close any vacuum system isolation valves to prevent suck-back into the2.
vacuum system (where installed).
3. For the 1-phase pump; stop the pump system using the appropriate3.
control source, that is, using the Stop button in manual control mode
r
efer to
; the Start/Stop control input in
parallel control mode or a Stop command in serial control mode.
For the 3-phase pump system; remove power by isolating the incoming
3-phase electrical supply.
4. Vent the pump system using the gas ballast control or the valve on the4.
inlet.
5. Isolate the Mains supply.5.
7.10Emergency Stop (EMS) (3-phase pump only)
CAUTION: PUMP DAMAGE
Risk of damage to the pump. The 3-phase pump is not tted with an emergency
stop. If an emergency stop function is required, then this product can be
connected into the user’s own Emergency Stop Control Circuit.