EasyManua.ls Logo

LEYBOLD SCROLLVAC plus - Maintenance Plan; Inspect and Clean the Inlet Strainer

LEYBOLD SCROLLVAC plus
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The pump is designed to require little user maintenance. Observe the
following guidelines when carrying out maintenance on the pump:
Make sure the maintenance is done by a suitably trained and supervised
technician. Obey local and national safety requirements.
Make sure the maintenance technician is familiar with the safety
procedures which relate to the products processed by the pumping
system.
Check that all the required parts are available and are of the correct type
before starting work.
9.2Maintenance plan
More frequent maintenance may be required if the pump is used to pump
aggressive gases or vapours, such as solvents, organic substances and
acids, or if the pump is operated continuously at the higher end of its
operating temperature.
Table 18 Maintenance plan
Operation
Frequency
(months)
Service
indicator
1-phase
Hours
indicators
1-phase
3-phase
Section
reference
Inspect and clean the
inlet strainer
12 No 8760

 on page 57
Inspect and clean the
external fan cover if
required
12 No 8760

 on page 58
Check the pump
performance
30 Yes 21900

 on page 58
Replace the tip-seal
30 or when
needed
Yes 21900
 on
page 58
Replace the pump
bearings
60 Yes 43800

 on page 58
Replace the pump
controller
120 Yes 87600


 on page 59
Electrical safety check 60 No 43800
 on
page 60



9.3Inspect and clean the inlet strainer
Whenever the pump is disconnected from the vacuum system, or on an
annual basis, we recommend:
Removing the inlet strainer from the pump inlet and remove any debris
that may have accumulated. Refer to


.
Inspecting the inlet strainer and if necessary, clean it with a cleaning
solution suitable for the substances pumped. Ret the inlet strainer
before reconnecting the pump to the vacuum system. Refer to

 on page 28.
300668736_002_C7 - © Leybold 57
Maintenance

Table of Contents

Related product manuals