11Storage
Use the following procedure to store the pump:
1. Shut down the pump as described in
 on page 42.1.
2. Disconnect the pump from the electrical supply.2.
3. Place and secure protective covers over the inlet and outlet ports.3.
4. Non-condensing conditions shall be assured to avoid condensation4.
from quick temperature changes during storage and transport.
5. Store the pump in cool, dry conditions until required for use. When5.
required, prepare and install the pump as described in
 on
page 26.
If the storage period is less than 1 year, pump can be used in a standard
way
If the storage period is between 1 and 2 years, we recommend to do
the following steps every half of the year:
Remove the black plastic plug from the motor stator, manually
rotate the plug few times by 4 mm Allen key in the direction
shown to ensure it is rotating free and allow the bearing to be
lubricated.
Remove Allen key and place the plug back.
Switch on the pump and run the pump at ultimate pressure for
10hours, then open GB to level1 and run the pump for another
14hours to clean and dry the pump.
We recommend to use exhaust lter (silencer).
If the storage period is longer than 2 years, we recommend you send the
pump to our service for cleaning and check pump condition with full testing.
300668736_002_C7 - © Leybold 67