LI-600 Porometer/Fluorometer Instruction Manual
4-4
Code Generator, the drop-down menu offers access to
the following:
Instrument Calibration shows user calibration coef-
ficients and user-adjustable parameters. Calibration para-
meters can be adjusted for the relative humidity (RH)
sensors, flow sensors, and a fluorescence detector if
there is a fluorometer. See Calibrating the sensors on
page8-9 and User calibration procedures on page9-10 for
instructions.
Factory Calibration appears in the drop-down menu
when Allow Factory Calibration is checked in Prefer-
ences. Fluorometer, Intensity, Flow In, Flow Out, IRT,
RH, and Miscellaneous values can be edited. Perform a
factory reset, download a calibrations file from the LI-
COR website, or perform a manual reset using a down-
loaded file. Select RH in the left menu to change the
RHsensor serial number and to calibrate.
Preferences: select the text delimiter for data and log
files (comma or tab). Check Allow Factory Cal-
ibration to enable Factory Calibration.
About Instrument displays the instrument serial num-
ber, instrument firmware version, instrument time,
porometer serial number, fluorometer serial number,
and the RH sensor serial number.
Buttons
Home: Goes back to the home screen.
Table 4-1. Toolbar buttons.