Section 7. Troubleshooting
7-5
The RH values and sample RH are high
If the RH values on the display are 95% or above, water
may have entered the cuvette (see External Cleaning on
page8-1). If the sample RH (RH Sample) value on the
display is 95% or above, clamp onto a stable object, such
as a piece of printer paper, and flush the instrument’s
system until the sample RH value drops to match the
reference RH (RH Ref).
Instrument warnings
This section describes the warnings that can appear on
the LI-600 display and how to address them.
Instrument warnings will appear with this
symbol on the instrument display.
To determine the warning type, open Instrument Set-
tings by pressing the left arrow from the Select Con-
figuration screen, then select Warnings.
Matching. Do not clamp on leaf.
Matching happens automatically when the LI-600 is
powered on, and when you select Match on the Instru-
ment Settings screen. Always allow the instrument to
match before clamping onto a leaf.
Low memory
The Low Memory warning appears when there is less
than 10% of data storage space remaining. Stop using
the instrument; download the data, then clear all data
from the instrument.