7-1
Section 7.
Troubleshooting
This section describes how to identify possible issues with
the LI-600 and any warnings that may appear on the instru-
ment display. If you cannot find a solution in this section,
contact your local distributor or LI-COR support (see Sup-
port on page1-4).
Instrument issues
This section describes the potential issues with the LI-600
and how to address them.
Instrument will not power on
If the instrument will not power, ensure that your battery
is charged and the power supply is adequate when char-
ging; the power supply needs to source at least 1 Amp at 5
VDC.
Ensure that your cable and battery are functioning prop-
erly (see Removing and inserting a battery on page8-7).
Instrument will not seal on a leaf
If the instrument will not seal on a leaf when taking a
measurement, re-clamp on the leaf while avoiding the
veins. If there is debris in the aperture clamp or leaf gas-
ket, carefully remove it with a foam swab then re-clamp.