Section 8. Maintenance
8-5
cloth, and isopropyl alcohol (IPA) to carefully clean the
fluorometer optics.
Internal Cleaning
Internal maintenance procedures require removal of the
instrument head shells.
Removing the instrument head shells
Work indoors in a clean environment when removing
the instrument shells.
1 Using a Phillips screwdriver, remove the seven screws on the
left side of the instrument head.
Figure 8-5. Remove the shell on the left side.
2 Remove the five screws on the right of the instrument. Set the
shells and screws aside for safekeeping.