EasyManua.ls Logo

LI-COR LI-600 - Page 96

LI-COR LI-600
143 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 8. Maintenance
8-17
1 Prepare the LI-600.
Power on the instrument.
2 Open Instrument Calibration.
Connect the LI-600 to a computer via USB then open
the computer software. Click the drop-down menu
on the Toolbar then select Instrument Calibration.
3 Set the H
2
O input.
Select RH Span then click Begin RHSpan Cal-
ibration to view the RH Ref and RH Sample. Select
either Dew Point or Mole Fraction then enter the
desired H
2
O input.
4 Power off the pump.
Select Utility then click TURNPUMP OFF.
5 Remove the leaf clamp.
Gently but firmly pull on the leaf clamp to remove it
from the instrument.
Figure 8-16. Leaf clamps.
6 Attach the calibration clamp.

Table of Contents

Other manuals for LI-COR LI-600

Related product manuals