LI-600 Porometer/Fluorometer Instruction Manual
8-20
serial number, the LI-600 will continue logging data
using the old sensor serial number.
Take the following steps to change the RH sensors serial
number.
1 Open Factory Calibration.
Connect the LI-600 to a computer via USB then open
the computer software. In Explore the LI-600, select
the drop-down menu on the Toolbar then select
About Instrument. Check the Factory Calibration
box then click Close. Select the drop-down menu
again to allow the software to register the change
then click Factory Calibration.
2 Calibrate the RH sensor.
Select RH under Factory Calibration then click Cal-
ibrate. Click OK on the confirmation prompt.
3 Change the RH sensor serial number.
The Change RH Sensor Serial Number and Auto-
matic RH Sensor Calibration buttons will appear.
Click Change RH Sensor Serial Number then
input the serial number. Click OK on the con-
firmation prompt.
4 Calibrate the RH sensor.
Click Automatic RHSensor Calibration to cal-
ibrate the RHsensor.