ApexRp Operators Manual
248083480-1 Rev 3 38
5 Maintenance Procedures
Introduction
This chapter provides routine maintenance instructions that the
ApexRp instrument requires. The maintenance procedures described
in this chapter are not required on regular or prescribed intervals and
should be performed only if the user has reason to question the data
they are receiving.
Safety
Before performing any of the maintenance tasks described in this
chapter, read Chapter 1 of this manual and become familiar with the
warnings and caution labels.
Maintenance Calibration
To maintain optimum performance of this instrument, it should be
recalibrated annually by a Lighthouse Authorized Service Provider.
Zero Count Test
This section will provide the user a procedure to determine if the
ApexRp can successfully complete several zero counts. A purge
filter must be attached to the instrument and six (6) five (5) minute
samples must be taken. There should be no more than 1 count on
average per five-minute sample.
1. Connect the Purge filter to the sample inlet.
2. Apply power to the instrument.
3. Configure the unit to sample for 30 minutes.
4. Allow the instrument to sample through a 30-minute period. This
time allows the unit to warm up and purge any residual particles
that might be inside it.
5. Configure the unit to sample for 5 minutes with a 10-second hold.
6. Allow the instrument to take 6 samples.
7. If an average of more than one count per five-minute period is
reported, allow the instrument to sample for 30 minutes to purge
it, then repeat the test (Steps 5 & 6).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lighthouse Worldwide Solutions ApexR3p and is the answer not in the manual?

Lighthouse Worldwide Solutions ApexR3p Specifications

General IconGeneral
BrandLighthouse Worldwide Solutions
ModelApexR3p
CategoryRemote Control
LanguageEnglish