Loading...
D9U001AM2-0101_0610
ACCUMULATOR INDICATOR MEANING
lit accumulator disconnected or faulty
ashing: 1,6s lit / 0,2s not lit accumulator deeply discharged
ashing: 0,1s lit / 0,1s not lit accumulator discharged
ashing: 0,2s lit / 1,6s not lit accumulator is charging
not lit accumulator charged
MAINS POWER LED MEANING
lit connected to the mains
ashing: 0,6s lit / 0,6s not lit keyboard error (ashing inverted to Lock LED)
error (rst fault)
ashing: 0,1s lit / 0,1s not lit service mode
not lit disconnected from the mains power
transformer switching error
1.8 Visual signalisation
VISUAL
SIGNALISATION
LOCK LED
lit
ashing: 
0,6s lit / 0,6s not lit not lit
Thighrest Lock LED locked lock
error
keyboard
error
motion
blocked
unlocked
Backrest Lock LED locked lock
error
keyboard
error
motion
blocked
unlocked
Bed Height, Trendelenburg and An-
titrendelenburg Tilt Lock LED
locked lock
error
keyboard
error
motion
blocked
unlocked
Foot Control Lock LED locked lock
error
keyboard
error
motion
blocked
unlocked
1.8.1 Mains Power LED (Attendant Control Panel)
1.8.2 Accumulator Indicator (Attendant Control Panel)
1.8.3 Lock LED (Attendant Control Panel)

Table of Contents