6 Use and Storage Conditions
DANGER!
Danger to life due to electric shock!
To ensure the bed’s class I protection against electric shocks:
► Ground the mains.
► Use exclusively Hospital Grade or Hospital Only receptacles for grounding.
CAUTION!
Minimal clearance underneath the bed (standard version with 15 cm castors) is 4,4 cm!
► Observe the path for any obstacles and avoid collisions and possible damages of any bed´s part on the underca-
rriage.
► Do not use bed lifts and hoists for lifting the bed (bed hoists for patients are permitted; clearance for the patient
hoists is 15 cm).
Multicare with OptiCare or with Symbioso are designed for use in rooms for medical purposes. Electrical installations must therefo-
re meet local norms laying down the necessary conditions for electrical installations.
► Disconnect the bed from the mains in exceptional cases (i.e. lightning or earthquake).
Multicare, OptiCare and Symbioso are not suitable for indoor environments:
► containing ammable gases (except oxygen cylinders).
NOTE There are applied no deviations to requirements of IEC 60601-1-2 ed. 4
NOTE There are no known other measures for keeping the basic safety based on EMC phenomena.
NOTE Beds equipped with integration module meet standard for IEEE 802.11 b/g/n (2400,0 MHz – 2483,5 MHz, modulation
DSSS (IEEE 802.11 b ), OFDM (IEEE 802.11 g/n) 20MHz bandwidth, EIRP = 0,34 W).