8.3 Potential Equalisation
The bed is equipped with a standard protective connector. This connector is used for potential equalisation between the bed and
any intravascular or intracardiac device connected to the patient to protect the patient from static electric shocks.
Fig. Potential equalisation connector – male
Fig. Potential equalisation connector -
female
Use equalisation connector if:
the patient is connected to any intravascular or intracardiac device.
Before connecting the patient to an intravascular/intracardiac device:
Connect the ground wire of the device to the potential equalisation connector (male) on the bed on which the patient in
question is lying.
Use a standard hospital connector (female).
Make sure that the connectors match.
Make sure that there is no possibility for inadvertent disconnection.
Before moving the bed:
Disconnect the patient from the intravascular or intracardiac device.
Disconnect the potential equalisation connector.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Multicare and is the answer not in the manual?

LINET Multicare Specifications

General IconGeneral
Height AdjustmentYes
ManufacturerLINET
ModelMulticare
Bed TypeElectric
Length ExtensionYes
CPR FunctionYes
Side RailsYes
External Width102.5 cm (with side rails)
External Length224 cm (standard)
Mattress Platform Width900 mm
Mattress Platform Length200 cm
Maximum Thighrest Angle30°
TypeHospital Bed
ApplicationHospital, Clinic, Home Care

Related product manuals