WM 100 TD and WM 100 TH EN 23
4.2 Hygienic preparation of the therapy device
4.2.3 Hygienic preparation during patient change
1. Wipe-disinfect the following parts:
Housing
Power supply cable
Power supply unit
2. Replace the following parts:
Breathing tube WM 24445
Respiratory mask with headband
Transport case prismaBAG basic WM 29659 / prismaBAG premium
WM 29977
Patient change set WM 29973
3. Open the therapy device (see "7.2 Opening the therapy device",
page 41).
4. Place the therapy device on a table on its rear.
5. Undo and remove 2 screws 23.
6. Undo and remove 2 screws 24.
7. Lift the central part of the housing 2 off the rear of the housing 1.
1
2
24
23