4.2 Hygienic preparation of the therapy device
4.2.3 Hygienic preparation during patient change
1. Wipe-disinfect the following parts:
• Housing
• Power supply cable
• Power supply unit
2. Replace the following parts:
• Breathing tube WM 24445
• Respiratory mask with headband
• Transport case prismaBAG basic WM 29659 / prismaBAG premium
WM 29977
• Patient change set WM 29973
3. Open the therapy device (see "7.2 Opening the therapy device",
page 41).
4. Place the therapy device on a table on its rear.
5. Undo and remove 2 screws 23.
6. Undo and remove 2 screws 24.
7. Lift the central part of the housing 2 off the rear of the housing 1.