4 Hygienic preparation
4 Hygienic preparation
4.1 General information
This product may contain disposable items. Disposable items
are intended to be used only once. Use these items only once
and do not reprocess them. Reprocessing disposable items may
impair the functionality and safety of the product and lead to
unforeseeable reactions as a result of aging, embrittlement, wear,
thermal load, the effects of chemical processes, etc.
Wear suitable protective equipment for disinfection work.
Please refer to the instructions for use supplied with the disinfectant
used.
Also observe the respective instructions for use of the device, the
components, and the accessories.
As an alternative to the disinfection procedure described here, you
can disinfect the device for a patient change up to 5 times using the
Keredusy procedure. Please follow the manufacturer's procedure
instructions. It is not necessary to exchange the components as per
the patient change set WM 29973 in this case. However, the air
filter WM 29651 and the pollen filter (if present) must be replaced.
4.2 Hygienic preparation of the therapy
device
4.2.1 Hygienic preparation during use
Please observe the instructions for use of the therapy device.
4.2.2 Hygienic preparation for repairs
1. Wipe-disinfect the exterior of the housing and the power supply
cable.
2. Clean or replace (depending on their condition) the following parts
in accordance with the instructions for use:
Breathing tube WM 24445
Respiratory mask with headband
3. Open the therapy device.
4. Clean points which are extremely dirty.
5. Change the filter (see "3.3 Changing the filter", page 20).
6. Close the therapy device (see "7.3 Closing the therapy device",
page 43).
Back
Forward
Contents
Previous page
Next page

Table of Contents

Other manuals for Lowenstein Medical prismaAQUA

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lowenstein Medical prismaAQUA and is the answer not in the manual?

Lowenstein Medical prismaAQUA Specifications

General IconGeneral
BrandLowenstein Medical
ModelprismaAQUA
CategoryHumidifier
LanguageEnglish