WM 67971b 09/2017
WM 100 TD and WM 100 TH EN 55
7.7 Replacing the parts of the housing
7.7 Replacing the parts of the housing
7.7.1 Replacing the front of the housing
1. Open the therapy device (see "7.2 Opening the therapy device",
page41).
2. Remove the mainboard (see "7.4.1 Removing the mainboard",
page44).
3. Remove the display (see "7.5.1 Removing the display", page49).
4. Press the On/Off button out of the front of the housing 3 from the
back.
5. Remove the seal for the system interface 42.
6. Detach and remove the connection cable for the hose heater 5.
(only applies to devices as of serial number 5.000).