70 EN WM 100 TD and WM 100 TH
WM 67971b 09/2017
12 Technical data
12.4 Pressure volume curve
pV curve at RV=0.5 l and f=20/min
Classification as per DIN EN 60601-1:
Operating mode Continuous operation
Electromagnetic compatibility (EMC)
acc. to DIN EN 60601-1-2:
Radio interference suppression
Radio interference immunity
Test parameters and limit values can be requested from the manufacturer if
required.
EN 55011 B
IEC 61000-4 Parts 2 to 6, Part 11, Part 8
IEC 61000-3 Parts 2 and 3
Warming of respiratory air Maximum +3°C
Maximum filling volume 400 ml
Pressure drop at a flow of (with 400 ml of water):
The pressure drop across the device combination of WM 100 TD therapy
device and WM 100 TH respiratory air humidifier does not increase.
Maximum flow 248 l/min
Maximum permitted operating pressure 40 hPa
Specification prismaAQUA
0
0.1
0.2
0.3
0.4
0.5
0.00 5.00 10.00 15.00 20.00
Volume
[l]
Pressure [hPa]
Back
Forward
Contents
Previous page
Next page

Table of Contents

Other manuals for Lowenstein Medical prismaAQUA

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lowenstein Medical prismaAQUA and is the answer not in the manual?

Lowenstein Medical prismaAQUA Specifications

General IconGeneral
BrandLowenstein Medical
ModelprismaAQUA
CategoryHumidifier
LanguageEnglish