WM 100 TD and WM 100 TH EN 65
12 Technical data
12 Technical data
12.1 Technical data of therapy device
Specification Therapy device
Product class according to 93/42/EEC IIa
Dimensions W x H x D in cm 17 x 13.5 x 18
Weight 1.4 kg
Temperature range
Operation
Storage
+5°C to +40°C
-25°C to +70°C
Permissible humidity during operation and storage Rel. humidity 15% to 93%, non-condensing
Air pressure range 700 hPa to 1060 hPa, corresponds to a height of 3000 m above sea level
Connection diameter of breathing tube (mask end)
in mm
19.5 (to fit standard cone)
Electrical power Maximum 40 VA
System interface
12 V DC
Maximum 10 VA
Current consumption during operation
230 V
115 V
in standby mode (Standby)
230 V
115 V
0.11 A
0.22 A
0.036 A
0.019 A
Classification as per DIN EN 60601-11:
Protection class against elec. shock
Degree of protection against elec. shock
Protection against harmful ingress of water and solid
particles
Protection class II
Type BF
IP21
Classification as per DIN EN 60601-1:
Operating mode
Continuous operation
Applied part Respiratory mask
Electromagnetic compatibility (EMC)
acc. to EN 60601-1-2
Radio interference suppression
Radio interference immunity
Test parameters and limit values can be requested from the manufacturer if required.
EN55011 B
IEC61000-4 Parts 2 to 6, Part 11, Part 8
IEC61000-3 Parts 2 and 3
Average sound pressure level in operation
as per ISO 60601-2-70
Approx. 26.5 dB(A) at 10 hPa (corresponds to a sound power level of 34.5 dB(A)
Average sound pressure level in operation
as per ISO 60601-2-70 with respiratory air humidifier
Approx. 27.5 dB(A) at 10 hPa (corresponds to a sound power level of 35.5 dB(A)
Sound pressure level of alarm beeper at least 58 db(A)
Alarms (optional)
All device types
Disconnection, severe leakage (optional)
prisma30ST, prisma30ST-C, prismaLAB
Apnea, low minute volume, low
Tidal volume
Alarms Disconnection, leak (optional)
Alarm output Optical and acoustic
CPAP operating pressure range 4 hPa to 20 hPa
AcSV pressure range 4 hPa to 30 hPa
BILevel pressure range 4 hPa to 25 hPa
Back
Forward
Contents
Previous page
Next page

Table of Contents

Other manuals for Lowenstein Medical prismaAQUA

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lowenstein Medical prismaAQUA and is the answer not in the manual?

Lowenstein Medical prismaAQUA Specifications

General IconGeneral
BrandLowenstein Medical
ModelprismaAQUA
CategoryHumidifier
LanguageEnglish