66 EN WM 100 TD and WM 100 TH
WM 67971b 09/2017
12 Technical data
Pressure accuracy
<
20 hPa: ± 0.6 hPa
>20 hPa: ± 0.8 hPa
P Lim
max
(maximum pressure in case of error) < 40 hPa
Target volume in AcSV mode
It is not possible to set a target volume for the AcSV mode. The pressure control
stabilizes the volume at the respective current level.
Automatic backup frequency in AcSV mode
The automatic backup frequency in AcSV mode is continuously adapted between
10bpm and 20 bpm depending on the filtered spontaneous rate and the relative
respiratory minute volume of the patient.
prisma25S-C
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Trigger
- Pressure rise rate
- Available modes
4 hPa to 25 hPa
4 hPa to 25 hPa
25% to 67%
Auto, can be set to 3 levels
Can be set to 3 levels
CPAP, S
prisma25S
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Trigger
- Pressure rise rate
- Available modes
4 hPa to 25 hPa
4 hPa to 25 hPa
25% to 67%
Auto, can be set to 3 levels
Can be set to 3 levels
CPAP, APAP, S, autoS
prisma25ST
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Trigger
- Pressure rise rate
- Backup frequency
- Available modes
4 hPa to 25 hPa
4 hPa to 25 hPa
25% to 67%
Auto, can be set to 3 levels
Can be set to 3 levels
0 bpm to 35 bpm
CPAP, APAP, S, autoS, autoS/T, S/T, T
prisma30ST
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Ti
- Trigger inspiration
- Trigger expiration
- Pressure rise rate
- Pressure drop rate
- Backup frequency
- Target volume
- Pressure adjustment
- Available modes
4 hPa to 30 hPa
4 hPa to 25 hPa
20% to 67%
500 ms to 4000 ms
Auto, can be set to 3 levels
Auto, can be set to 3 levels
Can be set to 4 levels
Can be set to 3 levels
Auto, 0 bpm to 35 bpm
300 ml to 2000 ml
Can be set to 3 levels
CPAP, APAP, autoS/T, S, S/T, T, aPCV
prisma30ST-C
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Ti
- Trigger inspiration
- Trigger expiration
- Pressure rise rate
- Backup frequency
- Available modes
4 hPa to 30 hPa
4 hPa to 25 hPa
20% to 67%
500 ms to 4000 ms
Auto, can be set to 3 levels
Auto, can be set to 3 levels
Can be set to 4 levels
0 bpm to 35 bpm
CPAP, S, S/T, T, aPCV
Specification Therapy device