5 Final test
5Final test
5.1 General information
The therapy device has a programmed step-by-step test to simplify the
test procedure. This is available when the Service mode is active
(see " Perform the step-by-step test", page 33).
The step-by-step test includes all the test steps which are necessary for
checking the proper functioning of the therapy device.
The test procedure is not dependent on the current settings of the
therapy device, i.e., the patient parameters are retained unaltered.
In addition, other service functions are available in Service mode (see
"5.2 Service mode", page 29).
• Perform a final test:
• Following every maintenance
• Following every repair/service
• Following every hygienic preparation
• During the final test, fill out the test record (see "13 Test record",
page 72).
• If you discover any defects or deviations from the specified values
during the final test, you must not use the therapy device again
until they have been rectified.
• You should also use the final test to locate faults. You will find an
overview of possible faults in the chapter "Error messages" (see "6
Faults", page 37).
• You will find the necessary auxiliary equipment in the chapter
"Required auxiliary equipment" (see "11 Required auxiliary
equipment", page 64).
5.2 Service mode
The therapy device has a Service mode, which can be activated with the
service connector WM 29917, which is available for the trained service
technician.
The Service mode is activated by connecting the service connector to
the system interface. The Service menu appears after approx. 5
seconds. The Service menu is only available in English and contains the
following functions: