12.1 Technical data of therapy device
Peak flow as per ISO 80601-2-70
CPAP and APAP mode
Test pressures:
4hPa
8hPa
12 hPa
16 hPa
20 hPa
AcSV mode
Test pressures:
4hPa
10.5 hPa
17 hPa
23.5 hPa
25 hPa
30.0 hPa
Pressure measured at the patient connection
opening at a flow of 40 l/min
4.0 hPa
8.0 hPa
11.9 hPa
15.9 hPa
19.9 hPa
4.0 hPa
10.4 hPa
17.0 hPa
23.5 hPa
25 hPa
30.0 hPa
Average flow at the patient connection
opening [l/min]
235 l/min
230 l/min
220 l/min
215 l/min
210 l/min
235 l/min
225 l/min
215 l/min
200 l/min
195 l/min
190 l/min
Warming of respiratory air Maximum +3°C
Stability of the dynamic pressure (short-term accuracy)
for 10 breaths/min as per ISO 17510-1:2007 when
using the 19 mm hose.
7hPa
10 hPa
13.5 hPa
20 hPa
Stability of the dynamic pressure (short-term accuracy)
for 15 breaths/min as per ISO 17510-1:2007 when
using the 19 mm hose.
7hPa
10 hPa
13.5 hPa
20 hPa
Stability of the dynamic pressure (short-term accuracy)
for 20 breaths/min as per ISO 17510-1:2007 when
using the 19 mm hose.
7hPa
10 hPa
13.5 hPa
20 hPa
Δp<0.24 hPa
Δp<0.28 hPa
Δp<0.3 hPa
Δp<0.4 hPa
Δp<0.24 hPa
Δp<0.32 hPa
Δp<0.4 hPa
Δp<0.48 hPa
Δp<0.4 hPa
Δp<0.32 hPa
Δp<0.46 hPa
Δp<0.56 hPa
Specification Therapy device