where the colposcope will be used is properly grounded.
6. If there are frequent disturbances in the power grid where the equipment is being used, prepare equipment
that will be able to stabilize the power voltage.
7.3 Battery
WARNING:
Replace the battery immediately if it is damaged or the capacity is exhausted.
Only use batteries that are provided by the manufacturer. Please contact your distributor or the
manufacturer for replacements.
Always turn the colposcope off first and disconnect the power cord from the colposcope as well as the AC
power source before replacing the battery.
Always disconnect the main cable from the colposcope before replacing the battery.
Battery is not required for the operation of The LT-300 SD and LT-300 HD digital video colposcopess. The
colposcopes are meant to be used with the power cord plugged into the wall socket. However, should the
necessity arise, the colposcope may be used on battery power.
NOTE: To use the colposcope with a TV or Computer, the colposcope must be plugged in to an AC outlet.
The battery is meant to stay inside the colposcope. Should the battery require replacement, the end user
must contact their distributor or the manufacturer for a replacement.
Battery Life Span
1000 charges or 5 years.
Work Duration
4.5 hours
Time for Full
Charge
3.5 hours
Charging Method
Automatic. As long as the colposcope is plugged into a working wall socket,
the batteries will recharge automatically.
Indicator Light
When the colposcope is plugged into the wall socket, the indicator light will be
green.
Type
Li-ion
Please reference the appendix for more battery details.
Tips on prolonging battery life:
Remove the battery if the colposcope will be placed into storage for longer than 30 days at a time.
The battery should be fully charged once every 30 days if the colposcope is not in use.
CAUTION: Should the battery fail to work or is damaged, it must be replaced immediately. Please contact
your colposcope representative or your hospital engineer for a replacement battery.
NOTE:
Please dispose of old batteries per local requirements.
Do not use or place the batteries near a heat source.
Do not reverse the positive and negative pole of battery during use.
Only authorized and trained personnel may remove the battery.
7.4 Storage
When the colposcope is not in use, cover the colposcope and store in a dry, clean and relatively dust free
area. Make sure the lens cap is always on the colposcope when it is not in use.
For long term storage, remove the battery and follow the above instructions. Then reference Appendix A,
Section II for specifications of the storage environment.