Appendix A Specification
NOTE: Unless specifically noted for one or the other, the following information is true for both the LT-300 SD
and the LT-300 HD digital video colposcopes.
I. Power Supply/Adapter
Adapter Model
UE36LCP1-150240SPA
Adapter Input
100-240V, 50/60Hz, 0.9A
Adapter Output
15V, 2.4A
Power Box Input Power
30VA
Power Box Output Power
15V, 1.7A
Colposcope Input Power
15V, 1.7A
II. Battery
Battery Life Span
1000 charges or 5 years.
Work Duration
4.5 hours
Time for Full
Charge
3.5 hours
Charging Method
Automatic. As long as the colposcope is plugged into a working wall
socket, the batteries will recharge automatically.
Indicator Light
When the colposcope is connected to an AC power source, the indicator
light is turned on. When the colposcope is not connected to an AC power
source and is using battery power, the indicator light is off.
Output
7.4V 2200mAh
Model Number
18650-2S
Type
Li-ion
III. Grounding
Ensure the equipment is grounded properly by plugging it into an AC receptacle with ground included.
Speak to the relevant department (such as building maintenance department) to ensure the AC receptacles
where the colposcope will be used is properly grounded.
IV. Environmental Specifications
The LT-300 SD and LT-300 HD digital video colposcopes Imaging System is designed for hospital use. Only
qualified and trained personnel may use the system. Make sure the environment under which the
colposcope operates is clean, properly ventilated, protected against shock, noise, and strong electrical
interference. In addition, the colposcope should not be placed under direct sunlight. The following are
environmental guidelines for the operation and storage of the colposcope.
Environmental
temperature
Operation
41°F~104°F (5
o
C
~40
o
C)
Storage
-4°F~131°F (-20
o
C
~+55
o
C)
Relative
humidity
Operation
≤85%
Storage
≤95%
Atmospheric
pressure
Operation
70.0kPa
106.0kPa
Storage
50.0kPa

Table of Contents