Electronic Fluxmeter EF 14
BA
no.: 100514EBA02
December 9
, 2020
Page
17
8. If you proceed with such a test, make sure that a suitable measuring range is selected. Use
the range key to set the right measuring range.
Caution!
The measuring coils must be treated with care. They may be fragile.
Important!
The test result with a field coil depends on the angle between the coil and the
magnetic potential lines. The greater the angle is between the coil and
magnetic potential lines, the greater will be the percentage error. For
example: A angle produces an error of 0.4%, a 10° angle produces an
error of 1,5%.