Description Command
('_' corresponds to a
space)
Command example
('_' corresponds to a
space)
Reply
('_' corresponds to a
space)
Permissible entry value Notes
Request measuring value ? ±XXX.XX
±XXX.XX_<
±XXX.XX_=
±XXX.XX_>
- -Reply dependent on measuring unit and resolution
-If tolerance mode is enabled, the current
tolerance state is indicated by the following symbols:
= : Within the tolerance
< : Below the tolerance
> : Above the tolerance
Request instrument designation ID? - C1200 - -
Request version number VER? - X.X - -
Change measuring unit to µm UM - UM - -
Change measuring unit to mm MM - MM - -
Change measuring unit to inch IN - IN - -
Change measuring direction to +A CHA+ - CHA+ - -
Change measuring direction to -A CHA- - CHA- - -
Switch device OFF OFF - OFF - -
Execute master measurement RST - RST - -
Enter factor FAC_<factor> FAC_2.5 FAC 0.0001 to 10.0000
--> max. 4 decimal places
-
Request factor FAC? - X.XXXX - -
Enter master MASTER_<master> MASTER_500 MASTER In mm = -999.990 to +999.990
--> max. 4 decimal places
In µm = -999990.0 to +999990.0
--> max. 1 decimal place
In inch = -39.369685 to +39.369685
--> max. 6 decimal places
-Reply dependent on measuring unit and resolution
Request master MASTER? ±XXX.XX - -Reply dependent on measuring unit and resolution
Enter nominal value NOMINAL_<nominal> NOMINAL_500 NOMINAL In mm = -999.990 to +999.990
--> max. 4 decimal places
In µm = -999990.0 to +999990.0
--> max. 1 decimal place
In inch = -39.369685 to +39.369685
--> max. 6 decimal places
-Reply dependent on measuring unit and resolution
Request nominal value NOMINAL? - ±XXX.XX - -Reply dependent on measuring unit and resolution
Deactivate tolerance mode TOL0 - TOL0 - -
Enable tolerance mode TOL1 - TOL1 - -
Enter higher and lower limit value TOL_<higher-
Tol>_<lowerTol>
TOL_300_-300 TOL <higherTol> or <lowerTol>:
In mm = -5.0000 to +5.0000
--> max. 4 decimal places
In µm = -5000.0 to +5000.0
--> max. 1 decimal place
In inch = -0.196850 to +0.196850
--> max. 6 decimal places
The input value for the <higherTol> must be greater than the
input value for the <lowerTol>
Request limit values TOL? - ±XXX.XX_±XXX.XX - -Reply dependent on measuring unit and resolution
Factory settings FSET - - - -Instrument must be switched on again manually
Request serial number SN? - YYMMXXXX - -
Deactivate serial number output SN0 - SN0 - -
Activate serial number output SN1 - SN1 - -
Enter resolution RES_<resolution> RES_3 RES Resolution = {1, 2, 3}
RES_1
RES_2
RES_3
mm
0.01
0.001
0.0001
µm
10
1
0.1
inch
0.0001
0.00001
0.000005
Request resolution RES? -
1
2
3
- -
Request calibration factor CCALF? - +/- XX.XXXXXX -
Set calibration factor CCALF_˂factor> CCALF 3.1234 CCALF 3.1234000
The customer calibration must be activated in the instrument
following input of the calibration factor CCALF
Every order or answer is completed with a carriage return (CR).
7.3 Data transmission parameters