3-14
slippage or device damage can occur.
NOTE: Do not disconnect the tubing while the system is on.
iii. Hang the saline bag on the saline pole at the back of the control module by moving the bag
around the right side of the control module to the back.
PRECAUTION: The saline spike should be below the saline bag when the bag is hung on the
saline pole, ensuring the spike tubing is not kinked. Separate the tubing lines
between the canister connector and the saline spike if more tubing is needed.
b. Connecting the Vacuum Tube Set to the Vacuum Canister
Connect the canister connector on the originating vacuum line to the patient port of the vacuum
canister lid by moving the connector around the left side of the control module to the lid.
Control Module’s flexible source tube to vacuum port (Step 5.c.)
Canister connector to
patient port
Saline spike to saline bag
(See Step 9.a., above.)