USER INSTRUCTIONS AND
OPERATIONS GUIDE
www.mammotome.com/ifu eIFU number: 001413

Table of Contents

Questions and Answers

  • H
    Hector PenaSep 23, 2025
    What does it mean if the mammotome REVOLVE Medical Equipment system reports: “Incompatible Probe – Replace Probe” during function execution with Error Code 451?
    • E
      ernest27Sep 23, 2025
      If the mammotome Medical Equipment system reports: “Incompatible Probe – Replace Probe” during function execution with Error Code 451, it means an incorrect type of probe is attached. Remove the probe attached to the holster and replace it with a new probe.
  • K
    kevin97Aug 27, 2025
    What should I do if the mammotome REVOLVE probe attachment is difficult?
    • T
      Tyler MccarthyAug 27, 2025
      If you're finding it difficult to attach the probe to the holster on your mammotome Medical Equipment, it might be because the probe is being loaded onto the holster at an angle. To resolve this, keep the probe parallel to the holster (along the long axis) when attaching it.
  • K
    Kirk BradyAug 28, 2025
    What to do if the mammotome Medical Equipment cutter remains stuck in the open aperture position?
    • M
      Michelle MillsAug 28, 2025
      If the cutter remains stuck in the open aperture position on your mammotome Medical Equipment, a homing error may have occurred. Remove the probe from the holster, rotate the cutter gear on the bottom of the probe until the aperture is only 1/3 open, then reattach the probe to the holster and reinitialize the system.
  • O
    orodriguezAug 29, 2025
    Why is there irregular gear noise during initialization of my mammotome Medical Equipment?
    • M
      Mark LopezAug 29, 2025
      Irregular gear noise during initialization of the mammotome Medical Equipment could be due to the probe not being correctly attached to the holster. Remove the probe and reattach it to the holster. If the noise occurs again, remove the probe and inspect the cutter gear on the probe. If it is not damaged, replace the probe on the holster and proceed through initialization. If the cutter gear is damaged, discard the damaged probe and continue with a new probe.
  • K
    Kara McfarlandSep 23, 2025
    What does it mean if the mammotome REVOLVE system reports: “Probe Error – Replace Probe” during function execution with Error Codes 450, 452, 454, 455, 456?
    • G
      Gail WangSep 23, 2025
      If the mammotome Medical Equipment system reports: “Probe Error – Replace Probe” during function execution with Error Codes 450, 452, 454, 455, 456, it means the system is unable to connect or communicate with the probe with the holster or that the biopsy count has been exceeded. Remove the probe attached to the holster and replace it with a new probe.
  • S
    Sara TateNov 24, 2025
    What to do if my mammotome Medical Equipment SMS collection chambers are not at the anticipated location after initialization?
    • J
      Jesus LaneNov 24, 2025
      If the Specimen Management System (SMS) collection chambers are not at the anticipated location after initialization of your mammotome Medical Equipment, it could be due to: * Probe/holster did not initialize correctly: Remove the probe and manually rotate the specimen collection chambers to align with the alignment indicator on the probe, being careful not to pinch or dislodge the chambers. Replace the probe on the holster and proceed through initializations. If Chamber 7 is desired for the next sample, select Advance to Chamber 7 on the touchscreen interface. * SMS cup was misaligned prior to attaching the probe to the holster: Remove the probe from the holster. Manually rotate the SMS cup chambers (not the outer cup) to align the 13th chamber with the sample port (marked by an ‘M’ ...
  • D
    Deborah Wilson DVMSep 12, 2025
    What should I do if the mammotome Medical Equipment system reports holster is not compatible: Error Code 203?
    • D
      Danny PittmanSep 12, 2025
      If the mammotome Medical Equipment system reports that the holster is not compatible and shows Error Code 203, it means that the control module and the holster have incompatible software versions. Use a holster that is compatible with the control module’s current software version.
  • P
    Phillip BakerSep 13, 2025
    What to do if the mammotome Medical Equipment system returns to Standby and reports error message: “Vacuum Error, Check Tubing Connections” Error Code 400 or 402?
    • K
      Kevin HaynesSep 13, 2025
      If the mammotome Medical Equipment system returns to Standby and reports error message: “Vacuum Error, Check Tubing Connections” Error Code 400 or 402, it could be due to a vacuum canister leak, or the vacuum pump is not working properly. Ensure the vacuum tubing is connected. Check that all vacuum canister lid ports are sealed and repeat system initialization.
  • S
    Stephanie ChenSep 6, 2025
    What to do if I cannot insert the valve cartridge into the control module’s vacuum connection slot in mammotome REVOLVE Medical Equipment?
    • T
      tammyorrSep 6, 2025
      If you cannot insert the valve cartridge into the control module’s vacuum connection slot in your mammotome Medical Equipment, it is likely due to misalignment of the valve slots. Manually align the valve cartridge T-slots with the orientation of the control module valve actuators. Alternatively, proceed through initialization without the valve cartridge attached to reset the valve alignment, and then reattach the valve cartridge to the control module.
  • M
    Maria BrewerSep 4, 2025
    What to do if fluid does not visibly move in the vacuum tubing when the Vacuum button is pressed and released on my mammotome REVOLVE Medical Equipment?
    • J
      Jessica SinghSep 4, 2025
      If fluid isn't visibly moving in the vacuum tubing when you press and release the Vacuum button on your mammotome Medical Equipment, it could be that the vacuum button is being held too long to trigger Clear/Dry Tap, but not long enough to trigger the Vacuum function. Adjust the length of time the user is pressing the button: faster for Clear/Dry Tap, longer for Vacuum.

Summary

Chapter 1 INTRODUCTION

Warnings and Safety Precautions

Highlights critical safety information, potential hazards, and warnings for safe operation.

General and System Safety Guidelines

Provides general usage precautions and specific safety guidelines related to the system's electrical and operational aspects.

Chapter 2 SYSTEM DESCRIPTION

Probe and Holster Types

Describes the different types of probes (ST, U/S, EX) and their corresponding holsters.

Chapter 3 INITIAL INSTALLATION AND GETTING STARTED

Unpacking and Component Assembly Procedures

Step-by-step instructions for unpacking system components and assembling them for initial use.

System Setup and Configuration

Guides through powering on the system, connecting components, and initial setup processes.

Chapter 4 STEREOTACTIC (ST) INSTRUCTIONS FOR USE

Chapter 5 ULTRASOUND (U;S) INSTRUCTIONS FOR USE

Chapter 6 EX INSTRUCTIONS FOR USE

Preparation and Steps for Tissue Sampling

Detailed instructions on how to prepare the system and perform tissue sampling using EX mode.

Biopsy Function and Operation Details

Explains the biopsy cycle, sample acquisition, and handling of collected tissue specimens.

Chapter 7 CLEANING AND DISINFECTION

Cleaning Control Module, Cart, and Footswitch

Detailed instructions for cleaning the main system components using specified detergents.

Cleaning and Disinfection for Holsters

Procedures for cleaning and disinfecting ST, U/S, and EX holsters using different options.

Chapter 8 SOFTWARE DESCRIPTION

Touchscreen Interface and Button Functions

Overview of the graphic user interface, touchscreens, and button functionalities on the control module.

Chapter 9 SERVICE AND TROUBLESHOOTING

Troubleshooting System Component Issues

Provides solutions for common problems related to system components like holsters and probes.

Troubleshooting Software and Error Codes

Addresses issues related to system software, error messages, and incompatible components.