Retracts the trocar shaft (needle).
• Press and hold ARM until blue LEDs flash
(about 1 second) to automatically retract the
trocar shaft to the pre-fire position.
Fires the trocar shaft (needle).
• Press and hold the mechanical safety unlock
button while simultaneously pressing the
FIRE button to advance the trocar shaft to
the post-fire position.
Detachable Components
The Bemis 800cc vacuum canister is a non-sterile, single-patient use device used with the Mammotome
revolve Dual Vacuum-Assisted Biopsy System to collect fluid from a biopsy procedure. Refer to Chapter 4 (ST
only), Chapter 5 (U/S only), or Chapter 6 (EX only) for instructions for use during a procedure based on
procedure type.
Figure 17. Vacuum Canister Detachable Component
The Mammotome revolve probe guide (ST only) is a sterile, single-patient use device used with a
Mammotome revolve ST probe during a stereotactic procedure. It is designed to guide and stabilize the probe
trocar shaft. Refer to Chapter 4: Stereotactic (ST) Instructions for Use for instructions for use during a
procedure.
Figure 18. Fork-Style Probe Guide
Figure 19. Button-Style Probe Guide