2-19
Table 6. Input Device Button Operations for ST Holster Only
Button
Description
ST Holster
ARM
Retracts the trocar shaft (needle).
Press and hold ARM until blue LEDs flash
(about 1 second) to automatically retract the
trocar shaft to the pre-fire position.
SAFETY
UNLOCK and
FIRE
Fires the trocar shaft (needle).
Press and hold the mechanical safety unlock
button while simultaneously pressing the
FIRE button to advance the trocar shaft to
the post-fire position.
Detachable Components
The Bemis 800cc vacuum canister is a non-sterile, single-patient use device used with the Mammotome
revolve Dual Vacuum-Assisted Biopsy System to collect fluid from a biopsy procedure. Refer to Chapter 4 (ST
only), Chapter 5 (U/S only), or Chapter 6 (EX only) for instructions for use during a procedure based on
procedure type.
Figure 17. Vacuum Canister Detachable Component
The Mammotome revolve probe guide (ST only) is a sterile, single-patient use device used with a
Mammotome revolve ST probe during a stereotactic procedure. It is designed to guide and stabilize the probe
trocar shaft. Refer to Chapter 4: Stereotactic (ST) Instructions for Use for instructions for use during a
procedure.
Figure 18. Fork-Style Probe Guide
Figure 19. Button-Style Probe Guide