sufficiently removed. Purified or deionized water should be used during final rinsing processes,
where applicable. (Multiple rinses may be required.) Refer to the manufacturer’s
recommendations for the removal of disinfectant residue.
NOTE: Disinfected products should be stored in a dry clean environment, protected from contamination,
direct sunlight, pests, and extreme temperature and humidity. Do not transport the device in a
container prior to disinfection.
The Mammotome revolve ST and U/S holsters have an expected life of 500 reprocessing cycles.
Cleaning the EX Holster
After each use, manually clean the EX holster thoroughly using one of the following wipes:
Active Ingredients
Concentration
Recommended
Product
Wet Contact
Time
Isopropanol
Ethylene Glycol Monobutyl Ether (2-
Butoxyethanol)
Diisobutylphenoxyethoxyethyldimethylbenzyl
ammonium chloride
17.2%
1%
0.28%
CaviWipes™ by
Metrex® Research
3 minutes
Isopropyl Alcohol
Alkyl (60% C14, 32% C16, 5% C12, 5% C18)
dimethyl benzyl ammonium chlorides
Alkyl (68% C12, 32% C14) dimethyl ethylbenzyl
ammonium chloride (0.25%)
55%
0.25%
0.25%
Super Sani-Cloth®
Germicidal
Disposable Wipe by
Nice-Pak/PDI, Inc.
2 minutes
Quaternary ammonium compounds, C12- 18-
alkyl [(ethylphenyl) methyl] dimethyl, chlorides
Benzyl-C12-18-alkyldimethyl ammonium
chlorides
0.14%
0.14%
Sani-Cloth® AF3
Germicidal
Disposable Wipe by
Nice-Pak/PDI, Inc.
3 minutes
1. Using a wipe specified above, wipe the holster to remove visible contamination.
2. Repeat Step 1 until no visible contamination is present. Additional wipes may be necessary to remove
all visible contamination.
Disinfecting the EX Holster
A low-level disinfection process must follow the cleaning procedure. The disinfectant wipes approved for use
with the EX holster are detailed in the “Cleaning the EX Holster” section.
1. Using a new wipe, thoroughly wipe the holster to maintain the specified contact time, per the wipe
manufacturer’s instructions. (See the table above.) Additional wipes may be necessary to maintain
the specified wet contact time.
2. Allow the holster to air dry. Do not expose the device to heat greater than 54°C.
NOTE: Disinfected products should be stored in a dry clean environment, protected from contamination,
direct sunlight, pests, and extreme temperature and humidity.
The Mammotome revolve EX holster has an expected life of 1,400 reprocessing cycles.
Cleaning the Holster Holder
After each use, manually clean the holster holder using a commercial wipe.