5-7
k. Continue the biopsy procedure, if desired.
Clearing a Clogged Probe
WARNING: The pressure differential created in the canister during the procedure can cause blood
discharge.
NOTE: If tissue, blood, or any other fluid clogs or fills the probe during the procedure, attempt to
clear the probe by pressing and immediately releasing the VAC button on the active input
device (holster, remote footswitch, or remote keypad) to initiate cutter movement and a
vacuum pulse.
NOTE: The specimen management system needs to be attached to the probe for this function to
be performed properly.
NOTE: Press and hold the VAC button to activate on-demand VACUUM for the duration of the
button press (up to 15 seconds). Quickly pressing and releasing the VAC button
commands the system to clear the probe.
Continue the biopsy procedure, if desired.
Disassembling the Mammotome revolve Biopsy System
When the tissue sampling procedure is complete, the user must discard the disposable system components.
Disconnecting the U/S Probe and Vacuum Tube Set
a. Disconnect the saline spike from the saline bag. Discard the saline bag using standard medical
technique.
b. Disconnect the vacuum tube set lines from the vacuum canister.
NOTE: Ensure the system is in standby before removing the vacuum tube set lines from the
vacuum canister. Do not disconnect tubing while system is at the Ready screen.
c. Disconnect the valve cartridge from the front of the control module.
d. Optional: Connect the probe’s two lateral vacuum tube set lines to each other and pinch the pinch
valve, if desired, for disposal.
NOTE: Do not remove vacuum tubing from the vacuum canister if vacuum is being used.
e. To remove the probe from the holster, press in on the locking tab on the side of the probe body.
Pull the probe backward about one inch and then to the right to disconnect it from the holster.
NOTE: Do not reinstall the probe’s protective sleeve. Doing so may cause injury to the user.