4-9
i. Rotate the stopcock so the OFF indicator is pointing toward the injection port.
PRECAUTION: Failure to have the stopcock in the OFF position may produce a partial tissue
sample or tissue sample of poor quality.
j. Press the VAC button for up to 15 seconds to clear fluids from the probe.
k. Continue the biopsy procedure, if desired.
Clearing a Clogged Probe
WARNING: The pressure differential created in the canister during the procedure can cause blood
discharge.
NOTE: If tissue, blood, or any other fluid clogs or fills the probe during the procedure, attempt to
clear the probe by pressing and immediately releasing the VAC button on the active input
device (holster, remote footswitch, or remote keypad) to initiate cutter movement and a
vacuum pulse.
NOTE: The specimen management system needs to be attached to the probe for this function to
be performed properly.
NOTE: Press and hold the VAC button to activate on-demand VACUUM for the duration of the
button press (up to 15 seconds). Quickly pressing and releasing the VAC button
commands the system to clear the probe.
Continue the biopsy procedure, if desired.
Disassembling the Mammotome revolve Dual Vacuum-Assisted Biopsy System
When the tissue sampling procedure is complete, the user must discard the disposable system components.
Disconnecting the ST Probe and Vacuum Tube Set
a. Using standard medical technique, remove the probe guide from the probe guide holder. Discard
the probe guide in the appropriate container after use.
b. Disconnect the saline spike on the vacuum tube set from the saline bag.
WARNING: Discard the saline bag using standard medical technique.
NOTE: Take care to ensure that no saline is spilled. Spills can damage the control module and/or
the keypad storage receptacle.