4-10
c. Disconnect the vacuum tube set line from the vacuum canister.
NOTE: Ensure the system is in standby before removing the vacuum tube set lines from the
vacuum canister. Do not disconnect tubing while system is at the Ready screen.
d. Disconnect the valve cartridge from the front of the control module.
e. Optional: Connect the probe’s two lateral vacuum tube set lines to each other and pinch the pinch
valve, if desired, for disposal.
PRECAUTION: Do not remove the vacuum tubing from the vacuum canister if vacuum is being
used.
f. To remove the probe from the holster, press the locking button to the unlocked position.
NOTE: Do not reinstall the probe’s protective sleeve. Doing so may cause injury to the user.
g. While maintaining the button in the unlocked position, slide the probe back and lift up.
The probe disconnects from the holster.
h. Discard the probe and vacuum tubing in the appropriate container(s) after use.
i. If a new probe is used to continue sampling, follow Steps 6-9 in Chapter 3: Initial Installation and
Getting Started, Setup to load a new probe.