Symptom Cause Remedy
Motor does not rotate, No Power. 1. Check power supply is
Display does not light. plugged into unit.
2. Check that power supply is
plugged into a live line.
3. Check connection of
power cord.
4. Check the line cord for
continuity and replace
if defective.
5. Return for servicing.
Motor does not rotate. Defective Remote Control or 1. Check that remote cable
Display lights. Setting Error. connector is fully inserted
into the receptacle.
2. Reapply power.
3. If motor still does not rotate,
select remote control in Main
Menu or Setup Menu and
verify settings.
4. Return to Mode screen and
verify icon shows.
Remote Control Mode.
5. See
Remote Control Mode
in
this manual for further details.